Laser Type:Carbon Dioxide Surgical Laser
Treatment Beam
Type:Sealed Carbon Dioxide, Liquid Cooled. Radio Frequency (RF) excited, 81.4 MHz
Active Medium:Low Pressure Mixture of Carbon Dioxide, Nitrogen, Helium, and Xenon Gases
Repetition Rate:1 to 5,000 Hz
Pulse Width:Up to 2 ms
US FDA CDRH Classification:Class IV
European MDD Laser Classification:Class 4
Principal Output (Invisible, Infrared):10.6 um
Beam Mode:TEM is Dominant Mode
Laser Beam Spot Size:Controlled by Delivery System
UltraPulse Mode Treatment Beam Settings
Energy-Per-Pulse:
1-mJ increments:2 – 10 mJ
5-mJ increments:10 – 30 mJ
10-mJ increments:30 – 100 mJ
25-mJ increments:100 – 225 mJ
Repetition Rate:1 to 1000 Pulses-Per-Second (Hz), Adjustable
Power:1 – 60 Watts, Adjustable
Timed Exposure:1 ms – 1.0 second, Adjustable in 1-ms Increments
Repeat Timed Exposure:On time = 0.01 – 1.0 second; Off time = 0.1 – 5.0 seconds
Continuous Wave (CW) Mode Treatment Beam Settings
Power:0.05-watt increments = 0.05-0.5 watts; 0.1-watt increments = 0.5-1.0 watts; 0.5-watt increments = 1.0 watts-5.0 watts; 1.0-watt increments = 5.0 watts-10 watts; 5.0-watt increments = 10 watts-60 watts
Timed Exposure:0.01 – 1 second, Adjustable in 1-ms increments
Repeat Timed Exposure:On time = 0.01 – 1 second; Off time – 0.1 – 5 seconds
Aiming Beam
Type:Diode
Power:5.0 mW maximum
US FDA CDRH Laser Classification:Class IIIa
European MDD Laser Classification:3A
Principal Output:630-650 nm
Cooling:Internal Closed Cycle Liquid-To-Air Heat Exchange
Height:100 cm (39 inches) to top of console; 119 cm (47 inches) to arm mount; 145-158 cm (57-60 inches) to top of vertical folded stored arm; 183-216 cm (72-85 inches) to top of vertical released arm; Width:34 cm (13.5 inches) front; 27 cm (10.5 inches) rear
Depth:51 cm (20 inches)
Weight:123 kg (270 lbs)